Apply to the requirements of GB/T2406, ISO4589 and ASTMD2863 (oxygen index method). Used to assess the combustion performance of the heterogeneity of solid materials, laminated materials, foam materials, films and membranes under the stated conditions.
Technical Characteristics
Tolerance heat glass tube: The diameter : "High 450 Φ75
With timer device: 999s±0.25s
With double Scale and pressure gauge device: the accuracy is ±5%(V/V)
The oxygen concentration tester can be matched (directly showed the test state of the oxygen concentration)
Mixed glass box
Exclusive clamp of U type
The diameter of ignition tube: Φ2±1mm, The length of the flame: 16±4mm